Kern-sohn VHE 2T5A Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Váhy Kern-sohn VHE 2T5A. KERN & SOHN VHE 2T5A User Manual [de] [en] [fr] Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 24
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating manual
Pallet truck scale
KERN VHE-A
Version 1.1
01/2014
GB
VHE-A-BA-e-1411
Zobrazit stránku 0
1 2 3 4 5 6 ... 23 24

Shrnutí obsahu

Strany 1 - Pallet truck scale

KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-

Strany 2 - KERN VHE-A

 Optimum accuracy in weighing results is achieved if the load centre is between the forks. The forks are easily bent and twisted when the load is

Strany 3 - 1 Technical data

7 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance

Strany 4

How to carry out adjustments  Switch off weighing system  Press for 30 sec., PAR is displayed  Press until CAL is displayed  Press , Z

Strany 5

 Press for approx. 3 sec., the count down for setting the second adjustment point runs off automatically. The weight for the second adjustment p

Strany 6

Carry out adjustment by entering the adjustment weight  Switch off weighing system  Press for 30 sec., PAR is displayed  Press until CA

Strany 7

 Acknowledge by .  Place the set weight and set the reference height  Press for approx. 3 sec., the count down runs off automatically, fo

Strany 8 - 5 Transport and storage

8 Commissioning Correct weighing only possible at „reference height“ (see fig. 1 and 2)  Actuate repeatedly the drawbar handle (hand lever 1 d

Strany 9 - 6 Unpacking and placing

8.1 Switch on the pallet truck scale  Switch on the pallet truck scale using key, the balance carries out a selftest and changes then to the zero

Strany 10

8.3 Weighing  Switch on the pallet truck scale using key, the balance carries out a selftest and changes then to the zero display „0“ is displa

Strany 11 - 7 Adjustment

8.4 Zero point correction  Switch on pallet truck scale using button.  Check whether the forks of the pallet truck scale are empty  Lift pal

Strany 12

GB KERN VHE-A Version 1.1 01/2014 Operating instructions Pallet truck scale Contents 1 Technical data ...

Strany 13

8.6 Totalization The display unit offers an option that allows the addition of weighing values and the display of the total weight.  Pick up the

Strany 14 - (example)

8.9 Battery operation Power is supplied via 2 x 1.5V AA batteries. At a regular use with full batteries, approx. 1 year can be weighed. If the bat

Strany 15

9 Error messages - O.L. - The weighing system is overloaded. Remove the load Err06 Load cell signal too high Err07 Load cell signal too low

Strany 16 - (see fig. 1 and 2)

10 Servicing, maintenance, disposal 10.1 Clean Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with m

Strany 17

11 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process mus

Strany 18

1 Technical data KERN VHE 2T5A Readability (d) 5 kg Weighing range (max) 2000 kg Reproducibility 20 kg Linearity ± 30 kg Stabilization time 3 s

Strany 19

1.1 Dimensions A Fork length 1150 mm B Fork width 150 mm C Minimum fork height 83 mm Clearance 30 mm D Maximum fork height 205 mm Hoisting he

Strany 20

2 Appliance overview 2.1 Overview of display Load lifted too high; upper arrow is flashing Reference height reached Reference height a

Strany 21

2.2 Keyboard overview 2.2.1 In weighing mode: Operation Short key pressing • Zeroing • Switch on balance • Add weight to sum memory l

Strany 22

3 Basic instructions for scales 3.1 Proper use The balance you purchased is intended to determine the weighing value of material to be weighed.

Strany 23

3.4 Monitoring of Test Resources In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testi

Strany 24 - 11 Instant help

6 Unpacking and placing 6.1 Site of installation, place of operation for scales The balances are designed in a way that reliable weighing results