KFB/KFN-TM-BA_IA-e-1220 41
Konfigurations-
daten
Display internal resolution
Position of the decimal dot
Readability [d] / verification value[s]
CAL
Adjustment, see chap. 6.9
Linearisation, see chap. 6.10
Balance with two weighing ranges and different maximum load
and weighing ranges and interval sizes but only one load-
supporting pan, whereby each range extends from zero to the
respective maximum capacity. When load is removed, weighing
scales will remain in 2nd range.
Display internal resolution
Position of the decimal dot
div.
div 1
Readability [d] / verification value [e]
1. weighing range
div 2
Readability [d] / verification value [e]
2. weighing range
CAP
CAP 1
Weighing scale capacity [max]
1. Weighing range
CAP 2
Weighing scale capacity [max]
2. Weighing range
CAL
Adjustment, see chap. 6.9
For linearization, see chap. 6.10
Weighing scales with one weighing range subdivided into partial
weighing ranges, each providing a different scale interval. The
scale interval depends on the applied load and is automatically
changed during loading and unloading.
Display internal resolution
Position of the decimal dot
div.
div 1
Readability [d] / verification value [e]
1. weighing range
div 2
Readability [d] / verification value [e]
2. weighing range
CAP
CAP 1
Weighing scale capacity [max]
1. Weighing range
CAP 2
Weighing scale capacity [max]
2. Weighing range
CAL
Linearisation, see chap. 6.10
s. Kap. 7.10 / 7.11
LOCK
ANM
Restricted taring
range
Press , the current setting will be displayed. Using the
navigation buttons (see chap. 2.1.1) select the desired setting, the
active digit is flashing.